					General Data | 
		
			| 
					Intrinsic error | 
					Refers to the reference conditions and is guaranteed for two years | 
		
			| 
					Warranty | 
					2 years | 
		
			| 
					Recalibration interval | 
					2 years recommended | 
		
			| 
					Quality system | 
					developed, designed, and manufactured according to DIN ISO 9001 | 
		
			| 
					Reference conditions | 
		
			| 
					  | 
					23 °C ± 2 K; 74 °F ± 2 K , Vm=230 V ± 10%, 50 Hz ± 0.1 Hz or 60 Hz ± 0.1 Hz | 
		
			| 
					Phase sequence: | 
					L1, L2, L3 | 
		
			| 
					Interval length: | 
					10 minutes | 
		
			| 
					Wye connection: | 
					(L1, L2, L3 to N) | 
		
			| 
					Power supply: | 
					88 V ... 265 V AC | 
		
			| 
					Environment conditions | 
		
			| 
					Working temperature range: | 
					-10 °C to 55°C; 14 °F to 131°F | 
		
			| 
					Operating temperature range: | 
					0 °C to 35°C; 32 °F to 95°F | 
		
			| 
					Storage temperature range: | 
					-20 °C to 60°C; -4 °F to 140°F | 
		
			| 
					Reference temperature range: | 
					23 °C ± 2 K; 74 °F ± 2 K | 
		
			| 
					Relative humidity: | 
					Fluke 1745: Class B2 acc. IEC 60654-1Fluke 1744/43: Class C2 acc. IEC 60654-1
 | 
		
			| 
					Housing | 
					robust, fully insulated housing and accessories | 
		
			| 
					Protection type | 
		
			| 
					Fluke 1745: | 
					IP50 as per EN 60529 | 
		
			| 
					Fluke 1744/1743: | 
					IP65 as per EN 60529 | 
		
			| 
					Safety | 
					IEC/EN 61010-1 600 V CAT III, 300 V CAT IV, pollution degree 2, double insulation | 
		
			| 
					Type test voltage | 
					5.2 kV rms, 50 Hz / 60 Hz, 5 s | 
		
			| 
					EMC | 
		
			| 
					Emission: | 
					IEC/EN 61326-1, EN55022 | 
		
			| 
					Immunity: | 
					IEC/EN 61326-1 | 
		
			| 
					Voltage and Current Measurement | 
		
			| 
					Input voltage | 
		
			| 
					Input range VI P-N: | 
					max 480 V AC | 
		
			| 
					Input range VI P-P: | 
					max 830 V AC | 
		
			| 
					Max. overload voltage: | 
					1.2 VI | 
		
			| 
					Input range selection: | 
					By job programming | 
		
			| 
					Connections: | 
					P-P or P-N, 1- or 3-phase | 
		
			| 
					Nominal voltage VN: | 
					<= 999 kV with PTs and ratio | 
		
			| 
					Input resistance: | 
					App. 820 kΩ per chan. Lx-N Single phase (L1 or A, L2 or B, L3 or C connected): app. 300 kΩ | 
		
			| 
					Intrinsic uncenrtainty: | 
					0.1 % of VI | 
		
			| 
					Voltage transformer: | 
					Ratio: <999 kV / VI | 
		
			| 
					Ration selection: | 
					By job programming | 
		
			| 
					Current input with Flexi Set | 
		
			| 
					Input ranges II L1 or A, L2 or B, L3 or C, N: | 
					15 A / 150 A / 1500 A / 3000 A ac | 
		
			| 
					Measuring range: | 
					0.75 A … 3000 A ac | 
		
			| 
					Position influence: | 
					Max. ± 2 % of measured value – for distance conductor to meas. head > 30 mm | 
		
			| 
					Stray field influence: | 
					< ± 2 A ac for Iext=500 A ac and distance to measuring head > 200 mm | 
		
			| 
					Temperature coefficient: | 
					< 0.05 % / K | 
		
			| 
					Current transformer: | 
					Ratio <= 999 kA / II | 
		
			| 
					Ratio selection: | 
					By job programming | 
		
			| 
					Connection: | 
					3-phase, 3-phase +N, 2 phase L1 or A and L3 or C (2 W-meter-method) 7 pole connector | 
		
			| 
					Current input for clamp | 
		
			| 
					Input ranges II L1 or A, L2 or B, L3 or C, N: | 
					0.5 V nominal (for II) 1.4 Vpeak | 
		
			| 
					Intrinsic uncertainty: | 
					< 0.3 % of II | 
		
			| 
					Max. overload: | 
					10 V ac | 
		
			| 
					Input resistance: | 
					App. 8.2 kΩ | 
		
			| 
					Current transformer: | 
					Ratio Ω 999 kA / II | 
		
			| 
					Power systems | 
					Delta, 2-Element Delta, Wye, Single Phase, Split Single Phase | 
		
			| 
					Logger | 
		
			| 
					Power supply | 
		
			| 
					Functional Range: | 
					88 V to 660 V absolute, 50 Hz / 60 Hz100 V to 350 V dc Internal fuse: 630 mA T
 | 
		
			| 
					Power consumption: | 
					5 Watts | 
		
			| 
					Ride through: | 
					Fluke 1745: Internal battery for typ. > 5 hours ride through with intelligent power managementFluke 1743/44: 3 sec Capacitor
 | 
		
			| 
					Fuse: | 
					Power supply fuse can be replaced in service facility only. Supply can be connected in parallel to measuring inputs (up to 660 V) | 
		
			| 
					Display, indicators | 
					LEDs for status and voltage levelsFluke 1745: LC-display with backlight for voltage, current, active power, phase sequence.
 | 
		
			| 
					Memory | 
		
			| 
					Intervals: | 
					Capacity 8 MB Flash-EPROMA function > 12000 intervals for > 85 days with 10 min intervals
 P function > 30000 intervals for > 212 days with 10 min intervals
 | 
		
			| 
					Events: | 
					> 13000 | 
		
			| 
					Memory model: | 
					linear, circular | 
		
			| 
					Interface | 
					RS 232, 9600...115 000 Baud, automatic Baud rate selection, 3-wire communication | 
		
			| 
					Dimensions | 
		
			| 
					Fluke 1745: | 
					282 mm x 216 mm x 74 mm (115in x 88in x 33in) | 
		
			| 
					Fluke 1743/1744: | 
					170 mm x 125 mm x 55 mm (69in x 51in x 22in) | 
		
			| 
					Weight | 
		
			| 
					Fluke 1745: | 
					approx. 3 kg | 
		
			| 
					Fluke 1743/1744: | 
					approx. 2 kg | 
		
			| 
					Measurement | 
		
			| 
					A/D converter: | 
					16 bit, sample rate: 10.24 kHz | 
		
			| 
					Anti-aliasing filter: | 
					FIR-Filter, fc =4.9 kHz | 
		
			| 
					Frequency response: | 
					Uncertainty < 1 % of Vm for 40 Hz to 2500 Hz | 
		
			| 
					Interval length: | 
					1, 3, 5, 10, 30 s, 1, 5, 10, 15, 60 minutes | 
		
			| 
					Averaging time for Min/max values: | 
					½, 1 mains period, 200 ms,1 , 3 , 5 s | 
		
			| 
					Time base: | 
					Resolution: 10 ms (at 50 Hz), deviation: 2s/day at 23 °C; 74 °F ± 2 K |