Basic Specifications
|
AC Withstand-voltage test portion
|
AC Voltage Output Component
|
Output voltage
|
0.2 to 5 kV AC (50/60 Hz), Single range output
Measurement category ? , Anticipated transient overvoltage 1500 V
|
Output frequency
|
50/60 Hz, ±0.2%
|
Voltage output method
|
PWM switching (0 V start, Voltage change during supply possible)
|
Voltage setting method
|
Digital setting, Setting resolution: 0.01 kV
|
Output voltage accuracy
|
±1.5% of the setting value ±20 V
|
Maximum rated power
|
100 VA AC (5 kV/20 mA), continuous
|
Transformer capacity
|
100 VA
|
Output waveform
|
Sine wave
|
Fluctuation rate
|
15% or less (At Maximum rated load → no load converge to set value within 1 second)
|
Distortion rate
|
5% or less (At 40 MΩ, load 5 kV output: impedance of measuring device during measurement)
|
Voltage measurement
|
True RMS indication
Digital meter, 0 to 5.00 kVrms AC Withstand-voltage, Accuracy ±1.5%rdg (±15 V for less than 1000 V)
|
Current Detection
|
Current measurement range
|
0.01 mA to 20 mA (2-range structure)
|
Designated value
|
True RMS indication
|
Measurement range
|
10 mA/20 mA
|
Measured area and resolution
|
0.00 mA to 9.99 mA, 0.01 mA (10-mA range)
0.0 mA to 20.0 mA, 0.1 mA (20-mA range)
|
Measurement accuracy
|
±2%rdg ±0.05 mA (10-mA range)
±2%rdg ±0.5 mA (20-mA range)
|
Decision Function
Decision method
|
Window comparator method (digital setting) for setting upper-limit and lowerlimit values
|
Decision contents
|
UPPER FAIL : when measured current exceeds the upper-limit setting
PASS : when measured current remains between the upper- / lower limit settings for set time
LOWER FAIL : when the measured current is below the lower-limit setting
UPPER LOWER FAIL : Test errors such as when there is no pre-set voltage out-put
|
Decision process
|
Output to the display, beeper sound, signals to EXT I/O for each decision result
|
Setting range
|
AC: 0.1 to 20.0 mA (Upper-limit value), 0.1 to 19.9 mA (Lower-limit value), Setting resolution: 0.1 mA
|
Insulation resistance test portion
|
Rated supply
|
500 V/1000 V DC (positive electrode)
|
No load voltage
|
1 to 1.2 times of rated voltage
|
Voltage measurement
|
Digital meter, 0 to 1000 V DC Insulation resistance, Accuracy ±30 V
|
Rated measurement current
|
1 to 1.2 mA
|
Short-circuit current
|
4 to 5 mA (500 V), 2 to 3 mA (1000 V)
|
Measurement range
|
0.2 M to 2000 MΩ (500 V), 0.5 M to 2000 MΩ (1000 V)
|
Accuracy guarantee
|
0.5 M to 999 MΩ (500 V), 1 MΩ to 999 MΩ (1000 V): ±4%rdg 1000 MΩ to 2000 MΩ: ±8%rd
|
Measured resolution
|
0.01 MΩ (0.20 MΩ to 19.99 MΩ)
0.1 MΩ (20.0 MΩ to 199.9 MΩ)
1 MΩ (200 MΩ to 2000 MΩ)
|
Measurement resistance range
|
2 MΩ, 20 MΩ, 200 MΩ, 2000 MΩ (500 V)
4 MΩ, 40 MΩ, 400 MΩ, 2000 MΩ (1000 V)
|